The EU Breakthrough Device (BtX) Designation Program
TL;DR
The EU’s upcoming BtX designation program aims to give manufacturers of novel, high-impact devices a faster, better-supported path to market without loosening safety standards:
- Two criteria determine eligibility for breakthrough status: A device must demonstrate high novelty in technology, procedure, or clinical application, and must show significant positive clinical impact or fill an unmet medical need for a life-threatening or irreversibly debilitating condition.
- Designation runs through an independent expert panel, not a notified body: Manufacturers request an opinion from EC-selected expert panels, who aim to issue a decision within 60 days and prioritize BtX reviews against their other workloads.
- Benefits extend across the full development lifecycle: Designated manufacturers gain early expert panel input, notified body prioritization during conformity assessment, and structured post-market support, mirroring the general intent of the FDA’s existing Breakthrough Devices Program.
What is the BtX Designation Program?
The European Medicines Agency (EMA) plans to launch a pilot programme for the EU Breakthrough Device (BtX) designation in the second quarter of 2026, as posted on the EMA Medical Device main page January 12, 2026. This pilot follows the approach laid out in the MDCG 2025-9 guidance document, released by the Medical Device Coordination Group (MDCG) in December 2025 entitled “Guidance on Breakthrough Devices (BtX) under Regulations 2017/745 & 2017/746.”
For manufacturers of novel devices that can make a significant impact to patient health, the goal of the program is to offer a path to streamlined and potentially faster market entry without sacrificing the rigour around ensuring safety and performance.
Breakthrough designation is meant for manufacturers of medical devices regulated under MDR 2017/745 and in vitro diagnostic devices (IVDs) regulated under IVDR 2017/746 that meet the defined criteria. This blog outlines the BtX criteria, how to obtain designation, the advantages of having BtX designation, and other considerations for manufacturers from the BtX guidance. In this discussion, the focus will be more on medical devices rather than IVDs.
BtX Criteria
For a medical device or IVD to be considered for breakthrough designation status, it must meet the following criteria, as defined by the MDCG guidance document.
- Novelty
The device introduces a high degree of novelty with respect to the device technology, the related clinical procedure, and/or the application of the device in clinical practice,
AND
- Positive clinical impact
The device is expected to provide a significant positive clinical impact on patients or public health, for a life-threatening or irreversibly debilitating disease or condition, by either of the following:- Offering a significant positive clinical impact on patients or public health compared to available alternatives and the state of the art,
OR - Fulfilling an unmet medical need where there is an absence or insufficiency of available alternative options for that purpose.
- Offering a significant positive clinical impact on patients or public health compared to available alternatives and the state of the art,
The guidance also explains how to interpret each criterion. Manufacturers are expected to describe the current state of the art and identify relevant alternatives for the same indication (including disease stage/subgroup) as well as justifying any exclusions using EU-appropriate standards and clinical guidance.
How to Obtain Designation
FTo receive BtX designation status, manufacturers must request an opinion from an expert panel. Expert panels are defined and established as per MDR 2017/745 Article 106. They are selected by the European Commission (EC) based on their scientific, clinical, and technical expertise. Expert panels provide independent opinions and scientific/technical advice on high-risk and orphan devices to manufacturers, notified bodies, member states, the MDCG, and the EC to help shape related guidance and standards.
The process and documentation requirements for seeking expert panel opinion have not been finalized. The EMA is expected to publish more detail before the pilot program starts. What is known is that manufacturers can make the request early in device development and, as per the BtX guidance, expert panels will aim to issue a BtX status opinion within 60 days and prioritize these reviews against their other workloads. Once a conclusion is reached, the Secretariat (administrative support for expert panels) notifies the MDCG. Devices that receive a positive opinion are then described as BtX-designated going forward.. Manufacturers are expected to describe the current state of the art and identify relevant alternatives for the same indication (including disease stage/subgroup) as well as justifying any exclusions using EU-appropriate standards and clinical guidance.
Advantages of BtX Designation for Manufacturers
BtX designation is intended to give manufacturers a more predictable and better-supported path from early development through certification and provide real-time feedback on critical path data gathering. The main advantage is the early alignment with expert panels, competent authorities, and notified bodies so that manufacturers can de-risk the clinical plan, carry out other key pre-market testing, avoid late cycle surprises, and keep review timelines moving. Below are some examples of benefits across the development lifecycle:
Pre-certification support during conformity assessment: notified body prioritization and tools like rolling interactions to reduce timeline uncertainty, potential use of the MDR expert panel pathway where applicable, and support through certification and ongoing surveillance activities
- Early support prior to and at designation: expert panel input on whether BtX criteria are met, plus regulatory support from national competent authorities (NCAs) once designation is confirmed
- Post-designation support through verification and validation and post-market planning: priority expert panel advice on clinical strategy and development plans, NCAs prioritizing clinical investigations, earlier and structured notified body interactions, and favourable cost considerations for small-to-medium enterprises
- Pre-certification support during conformity assessment: notified body prioritization and tools like rolling interactions to reduce timeline uncertainty, potential use of the MDR expert panel pathway where applicable, and support through certification and ongoing surveillance activities
Considerations for Manufacturers During Pre-Market Development
Pre-clinical
The guidance provides extensive coverage of considerations to BtX developers for pre-clinical testing and building technical documentation. The guidance touches on:
- Risk management file
- IVD development
- Medical devices containing artificial intelligence (AI)
- Biocompatibility
- Bench testing
- In silico modeling and simulation
- Pre-clinical literature research
- Usability
- Long-term non-clinical studies (e.g., shelf-life testing)
- Manufacturing process considerations
As an example, let’s look at BtX considerations for risk management. Due to the innovative nature of BtX devices, real-world data may be limited early in the risk management process. Risk acceptance criteria for new hazards should be based on what is currently considered acceptable in clinical practice, even if the likelihood of harm cannot be accurately estimated. As clinical results become available (through studies or in post-market), key assumptions and expected event rates should be checked and then updated over time.
Clinical
For pre-market clinical investigations, manufacturers should have a clinical evaluation plan (CEP) that clearly defines the intended patient benefit and outcomes, how safety and residual risks will be assessed, and how benefit–risk acceptability will be evaluated against current clinical practice. To support timely patient access, clinical investigations may focus on gathering sufficient safety and short-to-medium term performance data, with the expectation that this will be followed by post-market clinical investigations and other PMCF/PMPF (post-market clinical/performance follow-up) activities to confirm longer-term safety and performance.
Once on the market, manufacturers need to follow through on post-market commitments made to their notified body during conformity assessment, especially if their CE certificate includes specific conditions or provisions. These obligations may include PMCF/PMPFstudies within defined timelines and enhanced post-market surveillance and monitoring with updates to clinical and performance data. All of this is used to provide assurance and confirmation of the safety and performance of the BtX in real-world use throughout the device’s expected lifetime.
Moving Forward
For those familiar with the FDA’s Breakthrough Devices Program, the general intent of the EU version is similar; however, not surprisingly, the players and the framework around it are quite different. For more detail on the FDA, program refer to this resource.
On December 16, 2025, the EC published a legislative proposal to amend the MDR/IVDR for improvements related to reducing regulatory burden. Part of this amendment is to formally introduce the BtX process into the regulations and create a dedicated conformity assessment pathway. Implementation will depend on adoption through the European Parliament and Council. This is something that device developers should monitor on a regular basis.
In this complex world of bringing novel and clinically beneficial devices to market, it is great to see the EMA developing a framework to provide manufacturers with additional support. It is still early in the process and manufacturers, as well as key stakeholders such as the expert panels, notified bodies, and competent authorities will all need to become familiar with their respective roles in BtX designation. Manufacturers are looking forward to more information from the EMA and for Q2 2026, when the BtX pilot program is expected to start.
References
European Medicines Agency -> Medical Devices page, Announcement of Breakthrough Designation Pilot. https://www.ema.europa.eu/en/human-regulatory-overview/medical-devices
MDCG 2025-9 Guidance on Breakthrough Devices (BtX) under Regulations 2017/745 & 2017/746. Released December 2025. https://health.ec.europa.eu/document/download/edca94c7-62ab-4dd5-8539-2b347bd14809_en?filename=mdcg_2025-9.pdf
FDA Breakthrough Devices Program. https://www.fda.gov/medical-devices/how-study-and-market-your-device/breakthrough-devices-program
European Commission proposal to amend regulations (EU) 2017/745 and (EU) 2017/746 https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=comnat:COM_2025_1023_FIN
Deborah Pinchev is a StarFish Medical’s Toronto QA/RA Manager. She is inspired to write blogs by her interest in helping companies develop and bring to market new and innovative devices.
Images: Adobe Stock
Related Resources

Nigel Syrotuck breaks down REACH SVHC compliance for teams working with material suppliers and compliance questionnaires.

This medical vs wellness example shows how device classification can directly change functionality. Even when hardware is similar, what the device is allowed to do can be very different.

For manufacturers of novel devices that can make a significant impact to patient health, the goal of the program is to offer a path to streamlined and potentially faster market entry without sacrificing the rigour around ensuring safety and performance.

When I was starting out in medical devices, the discussion focused on the possibility of an internet of things and the promise of “big data” about everything.