FDA Guidance on Medical Device Cybersecurity


Accessibility Alt description: FDA sign in front of FDA building that says "U.S. Department of. health and Human Services Food and Drug Administration"
Resources

FDA Guidance on Medical Device Cybersecurity Overview

Authors: Russell Haley

TL;DR

FDA’s cybersecurity guidance requires manufacturers to treat security as a Total Product Life Cycle concern, not a one-time premarket check:

  • A Secure Product Development Framework (SPDF) formalizes cybersecurity risk management: identifying threats, documenting and evaluating risks, implementing and verifying mitigations, and monitoring for new threats become required inputs for every regulatory submission.
  • Device connectivity does not determine cybersecurity risk exposure: the FDA has rejected submissions over undocumented risks from features like USB ports or latent network ports, even on devices that don’t connect to the internet.
  • Cybersecurity documentation must be proactive rather than reactive: manufacturers cannot prove the absence of cybersecurity risk without first completing threat identification and risk analysis, making early SPDF implementation the only reliable path to compliance.

“Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” is an FDA Guidance on medical device cybersecurity that provides guidance on integrating cybersecurity into the quality system management and premarket submission process for medical devices.

It covers risk management, design controls, software validation, and other elements to ensure the safety, effectiveness, and security of medical devices in the face of potential cyber threats.

The goal of the FDA Guidance on medical device cybersecurity is to impart on device manufacturers the need to consider cybersecurity in all aspects of device software including design, development, testing, monitoring and maintenance. A key concept of the document is planning for the “Total Product Life Cycle”. The FDA has pulled in many aspects of cybersecurity that would normally be left to HIPAA and device customers. It is now incumbent on the device manufacturers to integrate secure software practices from the beginning of the development phase and show through documentation how they will continue to ensure the device remains secure.

What is Cybersecurity?

Figure 1: How cybersecurity relates to safety

NIST (National Institute of Standards and Technology) has an entire page dedicated to the various definitions of cybersecurity.

NIST identifies Cybersecurity as a synonym for computer security [1]. It is the “Measures and controls that ensure confidentiality, integrity, and availability of the information processed and stored by a computer.”[2] It is also “the prevention of damage to, unauthorized use of, exploitation of, and—if needed—the restoration of electronic information and communications systems, and the information they contain, in order to strengthen the confidentiality, integrity and availability of these systems.”[3]

When extended to medical devices, cybersecurity encompasses user and patient safety. As noted on the FDA Guidance on page 11: “The scope and objective of a security risk management process, in conjunction with other SPDF (Secure Product Development Framework) processes (e.g., security testing), is to expose how threats, through vulnerabilities, can manifest patient harm and other potential risks.” Figure 1 below shows the relationship between Cybersecurity Risk and Safety Risk.

How to manage Cybersecurity Risks

The FDA Guidance on medical device cybersecurity suggests that one way to manage cybersecurity risk is through what it calls a “Secure Product Development Framework” or SPDF. An SPDF is essentially a plan for identifying cybersecurity threats and mitigations for the entire lifetime of the device. Device manufacturers should implement an SPDF as part of the key Quality Management System (QMS) that govern how a device with software is developed and maintained.

The process can be boiled down to the following steps:

  • Identify threats
  • Document and evaluate risks
  • Document and implement mitigations
  • Test and verify mitigations
  • and finally monitor the device and device space for new threats.

The outputs of these steps and the implementation of future facing procedures are the inputs the FDA requires for all new regulatory submissions. While the steps of the process are easily identified, the implementation of an SPDF is anything but easy.

More About SPDF

A Secure Product Development Framework should become a standard part of any QMS used to develop an electronic medical device or any medical device with software. It is tempting to claim that a device does not require an SPDF or cybersecurity considerations, but this would be a mistake. There is no way to prove to the FDA that your device does NOT present cybersecurity risks until after you have performed many of the initial steps of an SPDF, namely Threat Identification and Cybersecurity Risk Analysis.

We have already heard stories of other manufacturers that assumed they were secure only to have the FDA point to a USB port or a latent network port (behind a panel no less!) and reject the application due to lack of cybersecurity documentation. Just because the device does not connect to the internet, does not mean that a threat actor couldn’t exploit a part of your system. The only way to prove that your device is safe is to implement an SPDF from the start of the device lifecycle and document the lack of threats or risks.

Cybersecurity Risk Management

Our software team works with a number of Cybersecurity professionals, Penetration Testing Firms and FDA consultants to ensure the devices we develop are able to meet the FDA Cybersecurity documentation requirements for medical devices. Device developers should internalize the FDA Guidance on medical device cybersecurity and create new processes to help with compliance and new tools for generating the outputs necessary to satisfy the FDA. This is not a small undertaking. We have spent considerable time and effort bringing these systems up. The effort has paid dividends for our customers because we have automated vulnerability detection and reporting.

If you would like to know more about how StarFish Medical can help with your cybersecurity regulatory compliance, please contact our Business Development group. They would be happy to discuss how StarFish can help you succeed in creating robust, secure medical devices.

[1] https://csrc.nist.gov/glossary/term/cybersecurity

[2] https://www.cnss.gov/CNSS/issuances/Instructions.cfm

[3] https://doi.org/10.6028/NIST.IR.8074v2

Image: FDA

Russell Haley is a StarFish Medical Senior Software Engineer. He is a software and IT veteran with over 20 years of start-up experience designing IoT systems that collect data through Wi-Fi, Bluetooth and MICS (implantable) radios and store vital records in the cloud from oceanographic buoys, financial institutions, passenger trains and most recently, medical device product design & development.

Businessman holding a glowing compliance icon with legal and regulatory symbols, representing REACH SVHC compliance for medical device manufacturers

Nigel Syrotuck breaks down REACH SVHC compliance for teams working with material suppliers and compliance questionnaires.

medical vs wellness device example showing alert vs no alert functionality

This medical vs wellness example shows how device classification can directly change functionality. Even when hardware is similar, what the device is allowed to do can be very different.

Engineers conducting pre-clinical testing of a novel medical device in a controlled laboratory environment

For manufacturers of novel devices that can make a significant impact to patient health, the goal of the program is to offer a path to streamlined and potentially faster market entry without sacrificing the rigour around ensuring safety and performance.

Medical device data management displayed on a connected healthcare tablet

When I was starting out in medical devices, the discussion focused on the possibility of an internet of things and the promise of “big data” about everything.

Explore Our Services

Illustration of a pencil and ruler

Our multidisciplinary teams have all the in-house capabilities to transform your medical device from concept to final product.

Illustration of a 3D cube

Medical device design and development with a commercial launch focus. See how we support clinical trials and facilitate seamless tech transfers.

Illustration of two speech bubbles

We help you navigate complex regulatory landscapes to achieve commercial success. Explore our integrated approach to quality and regulatory services.